100MHz, Dual Channel, Dual trace
CRT Readout Oscilloscope
S9000R Series

Features

- DC ¡­ 100 MHz frequency bandwidth
- 6-inch rectangular CRT with internal graticule
- High sensitivity: 1mV/div¡­5V/div
- Trigger level lock
- Hold-off time (S9040R, S9020R only)

 

- Panel setup:V/div.Vertical mode,X-Y,
   ALT/CHOP,UNCAL,ADD(SUB),x10,TV-V/H
- Cursor readouts :¡âV,¡âV%,¡âdB,¡âT,
   ¡â1/T,DUTY, PHASE
- Scale illumination
- Frequency counter up to 100MHZ

Specifications ============================================= S9000R Series
Specifications
S9100R
S9060R
S9040R
S9020R
CRT Configuration and useful screen
6-inch rectangular with internal graticule : 8 x10 div (1div =6mm)
Accelerating potential
Approx. +17KV (ref. Cathode)
Approx. + 2KV
Approx.1.5KV
Focusing
Possible
Trace rotation
Provided
Intensity control
Provided
Scale illumination
Provided

Vertical

System

Modes
CH1, CH2, DUAL, ADD
CH1, CH2, DUAL, ADD
Sensitivity
2mV/div ¡­ 5V ¡¾ 5% in calibrated 11 steps of 1-2-5 sequence
1mV/div ¡­ 5V/div ¡¾ 5% in calibrated 12 steps of 1-2-5 sequence
(1mV/div ¡­ 2mV/div ¡¾ 8%)
Variable ratio
¡Ã 2.5 : 1
Bandwidth (-3dB)
DC¡­100MHz
at 5mV/div
DC¡­20MHz
at 2mV/div
DC¡­60MHz at 5mV/div
DC¡­20MHz at 2mV/div
DC¡­40MHz
at 5mV¡­5V/div,
DC¡­15MHz
at 1mV¡­2mV/div,
DC¡­20MHz
at 5mV¡­5V/div,
DC¡­10MHz
at 1mV¡­2mV/div
Rise time
Approx.3.5ns
Approx. 6ns
Approx. 8ns,
Approx. 17.5ns
HF rejection
£¾20MHz
£¾20MHz
Transient response
Overshoot: 5%, Undershoot: 5% at 5mV/div
AC coupling
10Hz at -3dB
Impedance
Approx.1§Û ¡¾5% , 25pF
Approx.1§Û ¡¾2% , 25pF
Signal delay
provided
provided
Channel insulation
30 : 1
30 : 1
Maximum input voltage
400V (DC + AC peak), Freq : ¡Â 1KHz

Horizontal

System

Sweep mode
A, A intensity, B, B TRIG¡¯D
A
Sweep rate
A:0.5§Á/div¡­50ns/div ,
1-2-5 22 steps
B:50ms/div¡­50ns/div ,
1-2-5 19 steps
A:0.5s/div¡­0.1§Á/div ,
1-2-5 21 steps
B:50ms/div¡­0.1§Á/div,
1-2-5 18 steps
0.1§Á/div ¡­ 0.5s/iv, 1-2-5 21 steps
Variable ratio
¡Ã 2 : 1 (1-2step), ¡Ã 2.5 : 1 (2-5step)
¡Ã 2.5 : 1 continuously variable
Mag. ratio
x 10 ¡¾ 10%
 
Delay time jitter
Better than 1: 10000
 
Linear
x1 : ¡¾ 5%, x 10 : ¡¾ 15%
 
Sweep hold-off time
8 * Tstop, continuously variable
(Tstop : sweep stop time)
8 * Tstop, continuously variable
(Tstop : sweep stop time)

Trigger

System

Modes
Normal, Auto, Single time
Auto,Normal,Single time
Source
CH1, CH2, Line, External
CH1, CH2, Line, External
Coupling
AC, HF rejection,TV, DC (TV-V: 0.5s¡­0.1ms/div,
TV-H: 50§Á¡­0.1§Á/div)
AC, HF rejection,TV, DC (TV-V: 0.5s¡­
0.1ms/div, TV-H: 50§Á¡­0.1§Á/div)
Sensitivity normal
DC¡­20MHz: 1div,
20MHz¡­100MHz:1.5div
DC¡­20MHz: 1div,
20MHz¡­60MHz:1.5div
DC~40MHz
int:1.5div,ext.:
0.2div
DC¡­20MHz
int:1div,ext.:
0.2div
auto
30Hz¡­100Hz: 1.5div,
100Hz¡­20MHz:1div
20MHz¡­100MHz:1.5div
30Hz¡­100Hz: 1.5div,
100Hz¡­20MHz:1div
20MHz¡­60MHz:1.5div
30Hz¡­100Hz: 1.5div,
100Hz¡­20MHz:1div
20MHz¡­40MHz:1.5
div
30Hz¡­100Hz: 1.5div,
100Hz¡­20MHz:1div
TV
1.5 div
1.5 div
Ext. trig sensitivity
£¼20MHz : 50mVp-p , £¾20MHz : 150mVp-p
20MHz : 50mVp-p , £¾20MHz : 150mVp-p
TV external
150mV
150mV
Level lock & alt. trig
50Hz¡­100MHz
int.:2div,ext.:
0.25div
50Hz¡­60MHz
int.:2div,ext.:
0.25div
50Hz¡­40MHz
int.:2div,ext.:
0.25div
50Hz¡­20MHz
int.:2div,ext.:
0.25div

X-Y mode

Input
CH1 : X axis, CH2 : Y axis
CH1 : X axis, CH2 : Y axis
Sensitivity
Same as X, Y
Same as X, Y
X-axis bandwidth
DC ¡­ 2MHz at -3dB
DC ¡­ 1MHz at -3dB
DC ¡­ 2MHz at -3dB
Phase error
Less than 3¢ª (at DC ¡­ 50KHz)
¡Â3¢ª
(at DC¡­100KHz)
¡Â3¢ª (at DC¡­50KHz)
Z- axis System Bandwidth
DC ¡­ 5MHz
Max. input voltage
30V (DC + AC p-p), 1KHz or less
50V (DC + AC p-p), 1KHz or less
Impedance
Approx.20§Ú
Approx. 5§Ú

Frequency

Counter

Frequency range
20Hz ¡­ 100MHz
20Hz ¡­ 60MHz
 
Frequency
measurement
Freq. range
Display format
Resolution
20Hz¡Âf£¼100Hz
99.99Hz
0.1Hz
100Hz¡Âf£¼1KHz
999.0Hz
1Hz max
1KHz¡Âf£¼10K0Hz
9,990KHz
0.02KHz max
10KHz¡Âf£¼100KHz
99.99KHz

0.02KHz max

100KHz¡Âf£¼1MHz
999.9KHz
F £¼972kHz : 0.2KHz,
F ¡Ã972KHz: 2.6KHz 
1MHz¡Âf£¼10MHz
9.999MHz
3KHz max
10MHz¡Âf£¼100MHz
99.99MHz
30KHz max
 
Cursor Cursor readout

4digital, ¡âV, ¡âT, ¡â1/T

¡âV,¡âT,¡â1/T,¡â%,¡âdB,DUTY, PHASE 

Cursor

Measure-
Ment
Cursor readout 4digital,¡âV, ¡âT, ¡â1/T  ¡âV,¡âT,¡â1/T,¡â%,¡âdB,DUTY, PHASE 
Display mode ¡ä : variable, ¡å: base  ¡ä : variable, ¡å: base
Resolution 1 / 25 div  1 / 25 div
Range Vertical : ¡¾3 div, horizontal : ¡¾4 div  Vertical : ¡¾3 div, horizontal : ¡¾4 div
Panel setup V/div, vertical mode, CH2 inverter, ALT/CHOP,
UNCAL,ADD(SUB),x10, probe (x1, x10), 
X-Y,AT/D, TV-V, TV-H 
V/div, vertical mode, CH2 inverter,ALT/CHOP,
UNCAL,ADD(SUB),x10, probe (x1, x10), 
X-Y,AT/D, TV-V, TV-H 
Calibration

Signal
Waveform Rectangle
Amplitude 0.5 Vp-p ¡¾ 2 %  2 Vp-p ¡¾ 2 % 
Frequency 1KHz ¡¾ 2 %
Power Supply Line Voltage 90¡­250VAC ¡¾ 10 % 50Hz AC 110V/220V ¡¾ 10 %, 50Hz
Consumption Approx. 40W 
Environment ¡¡ Operating temperature : 0¡É ¡­ +40¡É,
humidity 90% RH (40¡É)
Storage temperature : -40¡É ¡­ + 60¡É, 
humidity 90%RH (50¡É)
Operating temperature : 0¡É ¡­ +40¡É,
humidity 85% RH (40¡É)
Storage temperature : -10¡É ¡­ + 60¡É, 
humidity 70%RH (50¡É)
Dimensions (W x H x D) mm
310 x 150 x 440
321 x 132 x 376
Weight
Net 6kg
Net 8kg
Accessory Standard
Operator¡¯s manual 1, power cord 1, probes 2,spare fuse 1